Photostability Chamber

Thermolab’s Photostability Chambers are designed in accordance with ICH 1Qb guidelines, monitoring temperature, humidity (optional) and light intensity.
Download Brochure
Standard Features Optional Features Specifications Documentation Technical Data
  • Double walled modular structure with 3” Thick PUF insulation
  • Inside Stainless steel 304 Mirror Finish & outside Stainless steel 304
  • Fluorescent Tubes for day light effect and UV tubes for ultra violet ray in accordance to the ICH guidelines requirement of a lighting energy for 1.2 mil lux hours and UV-A light for 200 watt hours/m2
  • Complete with standby refrigeration system in case the regular system fails
  • Inside observation glass door with foam type neoprene gasket
  • Data logging of UV and LUX Readings in the PLC along with date, time, temperature (and humidity if selected)
  • Automatic switch off of UV and fluorescent light when the test is over giving alarm.
  • UV light will be switched OFF when the door is open
  • 1 No. Stainless Steel perforated tray and 1 no. Light Bank
  • U shaped S.S. Nichrome wire air heaters
  • Compressor with CFC Free R 134 A / R 404 (Eco Friendly) refrigerant
  • PLC based control system with 7” colour HMI.
  • Safety controller sensor to cut off the supply in case of overshoot and undershoot of temperature giving audio visual alarm
  • Complete with IQ, OQ, PQ, Documents, Operations and Maintenance manuals
  • Works on 230 V AC single phase 50 Hz
  • 21 CFR Part 11 compliance window based communication software for data management
  • Humidity system
  • Stand by refrigeration system & Stand by humidity system
  • Scanner 4 point temperature scanner complete with sensors and printer interface and data storage facility. (4 point Humidity optional)
  • Door Access Security system
  • Mobile Alert system
  • Standard Model(S) : Inside S.S. 304 & outside mild steel powder coated
  • GMP Model(G) : Inside S.S. 304 & outside S.S. 304
  • Temperature Set Point : 25 °C or
    25 °C & 60% RH (Optional)
  • Temperature Accuracy : ± 0.2 °C
  • Temp Uniformity : ± 2 °C
  • Complete with IQ, OQ, PQ documents
  • Operation and Maintenance manuals
  • Training to end users
Models Capacity Internal Size External Size LIGHT BANKS No.s
(G) Ltrs. W x D x H cms W x D x H cms
TP 200 G 200 60 63 60 87.5 106.5 117 1